13
Document title
3.2.4
Setting the AwTherm Set Point Temperature
Figure 3-11 AwTherm Top Level Screen
With the Temperature Control parameter turned on (Section 3.2.3), and the lid (1) closed, the
AwTherm will control the temperature of the sample chamber environment (12). The heat pump
(6) controls the temperature of the air which is circulated through the airflow sleeve (7) around
the outside the walls of the sample chamber.
To set the temperature, from the Top Level screen (Figure 3-11), use the up and down arrows to
select the Temperature parameter. Press ENTER; the temperature parameter will flash (Figure 3-
12) and the label will change to SP (set point).
Figure 3-12 Select Temperature parameter
Use the UP/DOWN touch buttons to adjust the value to the desired temperature and press ENTER
to accept it.