9 - EN
EMC guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The G150 is intended for use in the electromagnetic environment specified below.
The customer or the user of the G150 should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The G150 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The G150 is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies build-
ings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The G150 is intended for use in the electromagnetic environment specified below.
The customer or the user of the G150 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment-guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic mate-
rial, the relative humidity should be at least
30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/output
lines
Not applicable
Not applicable
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Surge IEC
61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
Not applicable
Not applicable
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Guidance and manufacturer’s declaration-electromagnetic immunity
The G150 is intended for use in the electromagnetic environment specified below.
The customer or the user of the G150 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance
level
Electromagnetic environment-guidance
Voltage Dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
<5% UT(>95% dip in
UT) for 0,5 cycle
40% UT(60% dip in
UT) for 5 cycles
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a typi-
cal commercial or hospital environment. If
the user of the G150 requires continued op-
eration during power mains interruptions, it
is recommended that the G150 be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz) mag-netic
field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristics of a typical location
in a typical commercial or hospital environ
-
ment.