Ross Patrol Enteral Pump Operating Manual Download Page 6

CONTROL DIAL SETTINGS

DIAL

FUNCTION

7

OFF/CHARGE

Stops all pump functions. Battery charges in
all dial settings when pump is plugged in.
The pump will retain current 

RATE, DOSE

and 

VOL FED

values indefinitely unless the

pump is unplugged from an AC power
source while in the 

OFF/CHARGE 

setting.

OFF/

CHARGE

SET RATE

ml/hr

SET DOSE

ml

RUN

ml/hr

HOLD

VOL FED

ml

CLEAR

VOL

OFF/

CHARGE

SET RATE

ml/hr

SET DOSE

ml

RUN

ml/hr

HOLD

VOL FED

ml

CLEAR

VOL

125

OFF/

CHARGE

SET RATE

ml/hr

SET DOSE

ml

RUN

ml/hr

HOLD

VOL FED

ml

CLEAR

VOL

1500

OFF/

CHARGE

SET RATE

ml/hr

SET DOSE

ml

RUN

ml/hr

HOLD

VOL FED

ml

CLEAR

VOL

– 125

RUN

Only setting that activates feeding.  Visual
display shows rate of flow in mL/hr and the 
dash on the left side of the display scrolls from
top to bottom when pump is in 

RUN

mode.

NOTE:

It is normal for rotor movement to be

intermittent (eg, starts, stops, starts, stops, etc)
while in the 

RUN 

mode.  This intermittent move-

ment regulates the rate of flow, conserves bat-
tery life and does not affect the accuracy of
the pump.  The 

RUN 

mode should always

be verified by the top to bottom scrolling
bars on the left side of the display window.

SET DOSE

Allows dose to be limited (1 to 9999 mL in
1-mL increments) by pressing the keypad
arrows when the pump is in the 

SET DOSE

mode. The up arrow increases the dose.
The down arrow decreases the dose. 
Completion of this dose triggers the 

doSE

FEd

alarm.  Refer to 

doSE FEd

(page 13).

NOTE:

Pump will operate without a 

DOSE

selected (SET DOSE at 0000).

SET RATE

Allows flow rate to be set (1 to 300 mL/hr
in 1-mL/hr increments) by pressing the key-
pad arrows when the pump is in the 

SET

RATE 

mode.  The up arrow increases the

rate.  The down arrow decreases the rate.

NOTE:

The 

LOC 

feature may be activated

by depressing both arrows simultaneously
when the dial is in the 

SET RATE 

position.

Repeat procedure to deactivate.  The 

SET

RATE, SET DOSE,

and 

CLEAR VOL 

values

cannot be changed while the 

LOC 

feature

is activated.

52034/52036-M-21395-PTR.a  6/14/01  3:06 PM  Page 7

Summary of Contents for Patrol Enteral Pump

Page 1: ...s entyrapfoedjngtgbe Iqatallpymisptaocopdng tippuppsotlobelng TrrnpjmpdhaltfSPTRYTE Siljctfldwrdte 33o 305uL pr TusndiipltoSOTDISEand selhctdkse 1toi99omL if dosirod poti Ptmpwillip opuritewithoataDPSE selqctdd SKTDUSEat 0009 TsrndfialtoRhNtpstprt fepdlng Tkkcleprvpluuedmlivred tukhtoCJEKRVPL after fourbeTps vflumewjllbg cltarey SQE PPERATING MAPUAL FWRCOMPLKTEK KSTPUCTONS 0000 0000 000000 ROSS RO...

Page 2: ...el Displays 6 Control Dial Settings 7 8 Pump Sets 9 Instructions For Use 10 11 Checking Pump Accuracy 12 Troubleshooting 13 14 Cleaning 15 Storage 16 Service 16 Specifications 17 Warranty 18 1999 Ross Products Division Abbott Laboratories 52034 52036 M 21395 PTR a 6 14 01 3 06 PM Page 3 ...

Page 3: ...tion The Patrol Pump offers these features 1 Alarms for free flow 2 Flow rate selection from 1 to 300 mL hr in 1 mL hr increments 3 Accuracy to 10 or 0 5 mL hr whichever is greater refer to CHECKING PUMP ACCURACY page 12 4 Dose setting 5 Volume fed accumulation 6 8 hour battery operation at 125 mL hr when fully charged 7 Alarms FrEE FLO free flow no FLO no flow doSE FEd dose fed LO bAt low battery...

Page 4: ... times the hourly feeding rate should be hung NOTE Do not rely on free flow alarm as a criterion for use with volume sensitive patients NOT FOR IV USE Confirm proper placement and function of patient s enteral feeding tube nasogastric nasojejunal gastrostomy etc Failure to do so may result in vomiting and or aspiration Verify the following before initiating feeding 1 Only a Ross Patrol Pump Set is...

Page 5: ...ble eerposion hpzaad if ubdd in the proaence of foimmable anesthdaics WAURION Elrctrucal spick haraard Gf nzt ozeg cise Refwr servicpng to quplifled service pursotnel NARUING Diocolnect AP pewey coro buftre sprvifing Raplfce batkjery only with apjroved Rous rhpljcement pjgrt Troinding uelisbiliry wat wnry ba awhieved whkn this eduipmont is connrctew yo a iecoptacle marted Poskital Drede ITPUPT 143...

Page 6: ...ode This intermittent move ment regulates the rate of flow conserves bat tery life and does not affect the accuracy of the pump The RUN mode should always be verified by the top to bottom scrolling bars on the left side of the display window SET DOSE Allows dose to be limited 1 to 9999 mL in 1 mL increments by pressing the keypad arrows when the pump is in the SET DOSE mode The up arrow increases ...

Page 7: ... hr HOLD VOL FED ml CLEAR VOL 980 OFF CHARGE SET RATE ml hr SET DOSE ml RUN ml hr HOLD VOL FED ml CLEAR VOL 0000 CLEAR VOL Stops pumping action and resets the volume fed to zero NOTE The pump will beep four times before clearing the volume to alert users that the dial has been turned to CLEAR VOL The user must turn dial away from CLEAR VOL to any other setting before the fourth beep to avoid an un...

Page 8: ...l Pump offers a wide choice of feeding sets PUMP SET EXAMPLE SHOWN Ross Top Fill Enteral Nutrition Bag With Preattached Patrol Pump Set 9 pump insert blue connector Safe T Valve 52034 52036 M 21395 PTR a 6 14 01 3 06 PM Page 9 ...

Page 9: ...ube 9 Grasp the blue connector on the pump insert Loop the pump insert under the rotor counter clockwise 10 Gently pull up and then seat the blue connector in the slot above the rotor 7 Completely prime the set by allowing fluid to expel air from the tubing 8 Let fluid slowly enter sight cham ber to fill line 1 4 full NOTE If sight chamber is filled above the line after priming overfilled invert c...

Page 10: ...7 To see feeding volume delivered turn dial to VOL FED 16 Turn dial to RUN to start feeding 15 If desired turn dial to SET DOSE and select dose by pressing the arrows NOTE Pump will operate with out a DOSE selected SET DOSE at 0000 14 Select flow rate from 1 to 300 mL hr by pressing the arrows and observing the entry on the dis play Holding down an arrow will cause the numbers to scroll Never use ...

Page 11: ...pump according to instructions Suspend bag to side of or behind and 20 inches above pump see illustration With the test pump connected to AC power start the pump and let it run for 15 minutes Then run fluid into the graduated cylinder for 30 minutes and compare to SET RATE on pump Amount Expected in Graduated Cylinder SET RATE 30 minutes 76 mL hr 34 2 mL to 41 8 mL 150 mL hr 67 5 mL to 82 5 mL If ...

Page 12: ...tainer if necessary Be sure sight chamber is not overfilled or covered with a film of product Turn dial to RUN to start pump doSE FEd Dose complete VOL FED Turn dial to HOLD is equal to the SET DOSE Turn dial to SET DOSE and depress up arrow to increase dose OR depress down arrow to zero the dose setting OR turn dial to CLEAR VOL after four beeps volume fed will be cleared Turn dial to RUN to star...

Page 13: ... AC outlet for 12 hours then continue feeding NOTE After 30 minutes pump alarms continuously for one minute and then shuts down H5 HOLD countdown Turn dial to HOLD If desired program the HOLD timer for 1 90 minutes by depressing the up or down arrow LOC Lock out disables SET RATE Turn dial to SET RATE SET DOSE and CLEAR VOL Depress up and down dial settings arrows simultaneously to deactivate Repe...

Page 14: ...r exposure to tuberculosis use 70 concentration isopropyl alcohol These recommendations are not substitutes for official procedures that may differ among institutions DROP DETECTOR Clean the drop sensors with a soft brush soft lint free cloth or cotton swab and warm soapy water Then rinse the drop sensors with cotton swab and warm water Be sure no soapy film or residue is left on lenses Dry thorou...

Page 15: ...y occur In the event of a pump malfunction or for technical assistance or parts please contact Ross Products Division Abbott Laboratories OPERATING INSTRUCTIONS VIDEOTAPE An instructional videotape in VHS format assists caregivers and medical profes sionals in learning procedures for operating the Ross Patrol enteral delivery system This videotape is available for your use through a Ross represent...

Page 16: ...L hr Accuracy 10 or 0 5 mL hr whichever is greater refer to CHECKING PUMP ACCURACY page 12 Pressure Pumps against 18 psi nominal back pressure before no FLO alarm Refer to PATROL MAINTENANCE AND REPAIR MANUAL for test procedure Battery Operation Type Rechargeable sealed lead acid Voltage 6 V the pump will fully recharge in a maximum of 4 5 hours while pump is off or 12 hours if the pump is operati...

Page 17: ...in conjunction with the pur chase or use of the pump even if advised of the pos sibility of the same Ross reserves the right to repair or replace at its option any pump that fails to meet the foregoing warranty THE WARRANTIES HEREUNDER ARE IN LIEU OF ALL OTHER WARRANTIES EXPRESSED OR IMPLIED INCLUDING WITH OUT LIMITATION THE WARRANTIES OF MERCHANTABILITY AND OF FITNESS FOR A PARTICULAR PURPOSE NOT...

Page 18: ...21395 9901 52034 52036 10 99 LITHO IN USA 52034 52036 M 21395 PTR a 6 14 01 3 06 PM Page 20 ...

Reviews: