CONTROL DIAL SETTINGS
DIAL
FUNCTION
7
OFF/CHARGE
Stops all pump functions. Battery charges in
all dial settings when pump is plugged in.
The pump will retain current
RATE, DOSE
and
VOL FED
values indefinitely unless the
pump is unplugged from an AC power
source while in the
OFF/CHARGE
setting.
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
125
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
1500
OFF/
CHARGE
SET RATE
ml/hr
SET DOSE
ml
RUN
ml/hr
HOLD
VOL FED
ml
CLEAR
VOL
– 125
RUN
Only setting that activates feeding. Visual
display shows rate of flow in mL/hr and the
dash on the left side of the display scrolls from
top to bottom when pump is in
RUN
mode.
NOTE:
It is normal for rotor movement to be
intermittent (eg, starts, stops, starts, stops, etc)
while in the
RUN
mode. This intermittent move-
ment regulates the rate of flow, conserves bat-
tery life and does not affect the accuracy of
the pump. The
RUN
mode should always
be verified by the top to bottom scrolling
bars on the left side of the display window.
SET DOSE
Allows dose to be limited (1 to 9999 mL in
1-mL increments) by pressing the keypad
arrows when the pump is in the
SET DOSE
mode. The up arrow increases the dose.
The down arrow decreases the dose.
Completion of this dose triggers the
doSE
FEd
alarm. Refer to
doSE FEd
(page 13).
NOTE:
Pump will operate without a
DOSE
selected (SET DOSE at 0000).
SET RATE
Allows flow rate to be set (1 to 300 mL/hr
in 1-mL/hr increments) by pressing the key-
pad arrows when the pump is in the
SET
RATE
mode. The up arrow increases the
rate. The down arrow decreases the rate.
NOTE:
The
LOC
feature may be activated
by depressing both arrows simultaneously
when the dial is in the
SET RATE
position.
Repeat procedure to deactivate. The
SET
RATE, SET DOSE,
and
CLEAR VOL
values
cannot be changed while the
LOC
feature
is activated.
52034/52036-M-21395-PTR.a 6/14/01 3:06 PM Page 7