14
TROUBLESHOOTING
(continued)
VISUAL
CONDITION OR
EXPLANATION OR
DISPLAY
PROBLEM
CORRECTIVE ACTION
SEL run
HOLD
time-out
Turn dial to
RUN
.
E1 - E9
Electronics error
Remove pump from patient
immediately and contact Ross.
LO bAt
Approximately 30 minutes of
■
Turn dial to
HOLD
, then
RUN
.
battery operation remain
■
Plug power cord into AC outlet
for 12 hours, then continue
feeding.
NOTE:
After 30 minutes, pump
alarms continuously for one
minute and then shuts down.
H5
HOLD
countdown
■
Turn dial to
HOLD
.
■
If desired, program the HOLD
timer for 1-90 minutes by
depressing the up or down
arrow.
LOC
Lock-out disables
SET RATE,
■
Turn dial to
SET RATE
.
SET DOSE
and
CLEAR VOL
■
Depress up and down
dial settings
arrows simultaneously to
deactivate.
■
Repeat steps to activate.
8888
Display test
All segments should be
illuminated.