16
MODEL SOLU COMP II
SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS
1. SENSOR-OUTPUT ASSIGNMENTS (pH, ORP, or redox is selected during Quick Start)
6. MISCELLANEOUS SETTINGS
Section
Language
English
5.11
Hold
off
4.5
Security code
000 (no security code)
5.7
ac power frequency
60 Hz
5.8
5. TEMPERATURE RELATED SETTINGS
Section
Units
°C
5.6
Automatic temperature compensation
On
5.6
Solution temperature correction
Off
5.5
Isopotential pH
7.00
5.5
3. OUTPUT RANGES (pH, ORP, or redox is selected during Quick Start)
Measurement
Range
Section
pH
0 to 14
5.3
ORP/Redox
-1400 to 1400 mV
5.3
Temperature
0 to 100°C
5.3
Sensor(s)
Output 1
Output 2
Section
Single sensor
pH/ORP/Redox
Temperature
5.3 and 5.9
Dual sensor
pH/ORP/Redox for sensor 1
pH/ORP/Redox for sensor 2
5.3 and 5.9
2. OTHER OUTPUT SETTINGS
Output
Dampening
0 or 4 mA
Mode
Section
1
off
4
linear
5.3
2
off
4
linear
5.3
4. ALARM CONFIGURATION AND SETPOINTS
Alarm
1
2
3
Section
Assigned to
Sensor 1
Sensor 1 (note)
Fault
5.4
High or low
High
High
NA
5.4
Deadband
0
0
NA
5.4
Setpoint (pH)
14 (high); 0 (low)
14 (high); 0 (low)
NA
5.4
Setpoint (ORP/Redox)
1400 mV (high); -1400 mV (low)
1400 mV (high); -1400 mV (low)
NA
5.4
Note: For dual sensor input, alarm 2 is assigned to sensor 2.