Maintenance and Service
91
Reference Manual
Section 5: Maintenance and Service
00809-0100-4892
January 2019
FIGURE 5-10. DR Board Wiring, Right-Hand Side
9.
Refer to the wiring diagram in Figure 6-10. Reconnect the wiring to connector J4.
10. Connect the wiring harness plugs to connectors J1, J2, J3, and J8.
11. Slide the DR board completely into the 6888Xi enclosure.
12. Reapply power to the Remote Interface and Probe. When initialization is complete, the
Probe will begin heating up, but an alarm condition will be shown. Prior to using the
Transmitter for measurement purposes, perform one of the following:
a.
Recalibrate the O
2
Probe according to the applicable calibration instructions in Section 3.
b.
Synchronize the calibration parameters as follows:
i.
Press the <DIAG> button once to view the alarms. A “Probe Changed” alarm will be
shown.
ii.
Press the <EXIT> button once to enteer the “Diagnostic Menu”.
iii.
Use the Up/Down keys to select “3-Acknowledge Alarms”. Press <ENTER>.
iv.
Use the Up/Down keys to select “3-Ack Probe Changed”. Press <ENTER>.