Configuration, Startup and Operation
Section 3: Configuration, Startup and Operation
Reference Manual
January 2019
00809-0100-4892
TX
IOB
Parameter Name
Unit
Description
Y
Y
O2
%
Current oxygen concentration value (O2 %). The value should reflect
the last good O2 value if it is in the “Lock” state during calibration.
Y
Y
O2 Temp
°C
Current O2 sensor temperature.
Y
Y
CJC Temp
°C
Current cold junction temperature.
N
Y
Board Temp
°C
Current electronic temperature measured at the I/O board.
Y
Y
O2 Cell
mV
Raw mV value for O2 sensor.
N
Y
T/C Volt
mV
O2 T/C voltage.
Y
Y
Cell Imp
ohm
Cell impedance/sensor resistance measured.
Y
Y
Heater
V
Heater voltage.
N
Y
Bottle Pressure1
psi
Gas 1 bottle pressure. (Valid if device is configured for HSPS autocal.)
N
Y
Bottle Pressure2
psi
Gas 2 bottle pressure. (Valid if device is configured for HSPS autocal.)
N
Y
Cell Pressure1
in H2O
Cell pressure measured while applying gas 1. (Valid if device is
configured for HSPS autocal.)
N
Y
Cell Pressure2
in H2O
Cell pressure measured while applying gas 2. (Valid if device is
configured for HSPS autocal.)
Y
Y
O2 AO
mA
Analog output value represents the O2 concentration measurement.
N
Y
O2 AO %
%
Analog output percentage for O2 AO.
Y
Y
O2 Temp Max
°C
This is the highest O2 sensor temperature reached since last reset.
Y
Y
O2 Temp Max Time
Days ago
Time stamp of the highest O2 sensor temperature reached since
last reset.
Y
Y
Temperature | CJC Max
°C
This is the highest temperature reached at the cold junction since
last reset.
Y
Y
Temperature | CJC Max
Time
Days ago
Time stamp of the highest temperature reached at the cold junction
since last reset.
N
Y
Temperature | Board Max
°C
This is the highest temperature reached at the IO board since
last reset.
N
Y
Temperature | Board Max
Time
Days ago
Time stamp of the highest temperature reached at the IO board
since last reset.
Y
Y
Voltage | Heater Max
°C
This is the highest heater voltage reached since last reset.
Y
Y
Voltage | Heater Max
Days ago Time stamp of the highest heater voltage reached since last reset.
Y
Y
Voltage | Cell Max
°C
This is the highest cell voltage reached since last reset.
Y
Y
Voltage | Cell Max Time
Days ago Time stamp of the highest cell voltage reached since last reset.
N
Y
Ramp Rate | Heater Max
°C
This is the highest heater ramp rate reached since last reset.
N
Y
Ramp Rate | Heater Max
Days ago Time stamp of the highest heater ramp rate reached since last reset.
Y
Y
OP Mode
--
Device operating mode:
PO=Power up; WU=Warm Up (analog output is railed);
NM=Normal operation; CA=Calibrating (analog output can
be tracking or locked at last good value based on
"AO Tracks" configuration); AL=Alarm detected
(recoverable); SF=Alarm detected (non-recoverable)
Y
Y
Active Alarms
--
Current Alarms (See Section 4, Troubleshooting)
Y
Y
Device
--
Alarm state
Y
Y
Failed
--
Alarm state: On/Off
NV Memory Fail, Board Temp High, Factory Mode, Heater Ramp
Table 3-5. Probe Parameters (1 of 4)
Summary of Contents for 6888Xi
Page 2: ...B Table of Contents ...
Page 4: ...II Table of Contents ...
Page 8: ...VI Table of Contents Table of Contents Reference Manual January 2019 00809 0100 4892 ...
Page 32: ...24 Installation Section 2 Installation Reference Manual January 2019 00809 0100 4892 ...
Page 86: ...78 Troubleshooting Section 4 Troubleshooting Reference Manual January 2019 00809 0100 4892 ...
Page 113: ...105 ...