Configuration, Startup and Operation
69
Reference Manual
Section 3: Configuration, Startup and Operation
00809-0100-4892
January 2019
tion was known to contain invalid values. Resetting a calibration will transfer the slope and con-
stant from calibration 1 in the log into the calibration 0 position. The cell impedance and days
since calibration remain unchanged.
Reset the calibration as follows:
Using the 6888Xi Keypad & Display:
IOB 1 Probe 1(or 2)\Calibration\Cal Constants\
Current Cal\Reset Cal
Using the Field Communicator:
Configure\Calibration\Calibration Constants\
Reset Calibration Constants
Follow the instructions on the screen to complete the reset function.
3.9 D/A Trim
The D/A trim procedure is used to calibrate the 4-20 mA output signal to a precision mA meas-
urement device. Only the signal to the DCS needs to be trimmed.
D/A Trim with 6888Xi
1.
Press the MENU button repeatedly until the IOB1 Probe 1 menu displays.
2.
From the main menu, select DIAGNOSTICS then press ENTER.
3.
From the DIAGNOSTICS menu, select MAINTENANCE then press ENTER.
4.
From the MAINTENANCE menu, select ANALOG OUTPUT then press ENTER.
5.
From the ANALOG OUTPUT menu, select TRIM mA OUTPUT then press ENTER.
6.
In the first screen, a “Remove Loop from Auto” warning occurs. Remove the 6888Xi from
any automatic controls loops to avoid a potentially dangerous operating condition. Press
FIGURE 3-8. D/A Trim Methods, Simplified
WARNING
Failure to remove the 6888Xi from automatic control loops prior to performing this procedure may result
in a dangerous operating condition.
Summary of Contents for 6888Xi
Page 2: ...B Table of Contents ...
Page 4: ...II Table of Contents ...
Page 8: ...VI Table of Contents Table of Contents Reference Manual January 2019 00809 0100 4892 ...
Page 32: ...24 Installation Section 2 Installation Reference Manual January 2019 00809 0100 4892 ...
Page 86: ...78 Troubleshooting Section 4 Troubleshooting Reference Manual January 2019 00809 0100 4892 ...
Page 113: ...105 ...