MODEL 396/396VP/397/398/398VP
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
2
1.2 PERFORMANCE & PHYSICAL SPECIFICATIONS
SPECIFICATIONS
MODELS 396 &396VP MODELS 398 & 398VP
MODEL 397
QUIK-LOC KIT
Measurements
pH: 0-14
pH: 0-14
pH: 0-14
—
and Ranges
ORP: -1500 to 1500 mv
Available pH
GPHT hemi bulb or
GPHT hemi bulb or
GPHT hemi bulb
—
A
CCU
G
LASS
Types
GPLR flat bulb
GPLR flat bulb
Wetted Materials
316 SST, Polypropylene,
Titanium, Tefzel, glass,
Polypropylene, EP, glass
316 SST, EP, PEEK
EPDM, glass
choice of Kalrez, Viton, or
EPDM (platinum: ORP only)
Process Connection
None, use 1 in. NPT
None, use 1 in. NPT
None, must use Quik-Loc
1-in. MNPT
process connector,
process connector,
kit which includes 1 in.
PN 23166-00 or 23166-01
PN 23166-00 or 23166-01 MNPT process connection
(sold separately)
(sold separately)
Temperature Range
0-100°C (32-212°F)
0-100°C (32-212°F)
0-80°C (32-176°F)
—
Pressure Range-
100-1136 kPa (0-150 psig) 100-1825 kPa (0-250 psig)
100-790 kPa (0-100 psig)
—
Hemi bulb
Pressure Range-
100-790 kPa (0-100 psig)
100-790 kPa (0-100 psig)
—
—
Flat bulb
Minimum Conductivity
75 µS/cm, nominal
75 µS/cm, nominal
75 µS/cm, nominal
—
Preamplifier Options
Remote
Remote
Remote
—
Weight/Shipping Weight
0.45 kg/0.9 kg (1 lb/2 lb)
0.45 kg/0.9 kg (1 lb/2 lb)
0.45 kg/0.9 kg (1 lb/2 lb)
0.45 kg/0.9 kg (1 lb/2 lb)
PERCENT LINEARITY
396, 396VP, 397, 398, 398VP
396, 396VP, 397, 398, 398VP
pH Range
GPHT Hemi
GPLR Hemi
0-2 pH
94%
93%
2-12 pH
99%
98%
12-13 pH
97%
95%
13-14 pH
92%
—
Summary of Contents for 396