MODEL SOLU COMP II
SECTION 5.0
PROGRAMMING THE ANALYZER
TABLE 5-1. DEFAULT SETTINGS
1. SENSOR-OUTPUT ASSIGNMENTS
(type of dissolved oxygen measurement, pH, ORP, or redox is selected
during Quick Start)
3. OUTPUT RANGES
(type of dissolved oxygen measurement, pH, ORP, or redox is selected during Quick Start)
Measurement
Range
Section
Dissolved oxygen - ppm
0 to 20 ppm
5.3
Dissolved oxygen - ppb
0 to 900 ppb
5.3
Dissolved oxygen - % sat.
0 to 200%
5.3
pH
0 to 14
5.3
ORP/Redox
-1400 to 1400 mV
5.3
Temperature
0 to 100°C
5.3
Sensor(s)
Output 1
Output 2
Section
Single sensor
Dissolved oxygen
Temperature
5.3 and 5.9
Dual sensor
Dissolved oxygen (sensor 1)
pH/ORP/Redox (sensor 2)
5.3 and 5.9
2. OTHER OUTPUT SETTINGS
Output
Dampening
0 or 4 mA
Mode
Section
1
off
4
Linear
5.3
2
off
4
Linear
5.3
4. ALARM CONFIGURATION AND SETPOINTS
Alarm
1
2
3
Section
Assigned to
Sensor 1 (dissolved oxygen)
Sensor 2 (pH/ORP) (note)
Fault
5.4
High or low
High
High (note)
NA
5.4
Deadband
0
0
NA
5.4
Setpoint
0 ppm (high); 20 ppm (low)
0 ppm (high); 20 ppm (low)
NA
5.4
(Dissolved oxygen - ppm)
Setpoint
0 ppb (high); 990 ppb (low)
0 ppb (high); 990 ppb (low)
NA
5.4
(Dissolved oxygen - ppb)
Setpoint
0 ppm (high); 200% (low)
0 ppm (high); 200% (low)
NA
5.4
(Dissolved oxygen - % sat.)
Setpoint (pH)
14 (high); 0 (low)
14 (high); 0 (low)
NA
5.4
Setpoint (ORP/Redox)
1400 mV (high); -1400 (low)
1400 mV (high); -1400 (low)
NA
5.4
Note: For single sensor input, alarm 2 is assigned to sensor 1, and is configured as a low alarm.