MODEL SOLU COMP II
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
2
1.2 SPECIFICATIONS - General
Case: ABS. Pipe, surface, and panel mount versions
are NEMA 4X/CSA 4 (IP65).
POLLUTION DEGREE 4: Extended Environment
Outdoor use where conductive contamination
such as rain, snow, or dust may be present.
Dimensions
P
Pa
an
ne
ell ((c
co
od
de
e -1
10
0))::
6.10 x 6.10 x 3.72 in. (155 x
155 x 94.5 mm)
S
Su
urrffa
ac
ce
e//P
Piip
pe
e ((c
co
od
de
e -1
11
1))::
6.23 x 6.23 x 3.23 in.
(158 x 158 x 82 mm); see page 5 for dimensions
of pipe mounting bracket.
Conduit openings: Accepts PG13.5 or 1/2 in. con-
duit fittings
Display: Two line, 16-character, back-lit display.
Character height: 4.8 mm. Display can be cus-
tomized to meet individual requirements.
Depending on number of sensors, as many as 14
display screens are available.
Ambient temperature and humidity: 0 to 50°C,
(32 to 122°F) RH 5 to 95% (non-condensing)
Note: The analyzer is operable from -20 to 60°C
(-4 to 140°F) with some degradation in display
performance.
Power:
Code -01: 98 to 132 Vac, 50/60 Hz, 8 watts max.
196 to 264 Vac, 50/60 Hz, 8 watts max.
Code -02: 20 to 30 Vdc, 8 watts maximum
Equipment protected throughout by double insula-
tion.
RFI/EMI:
EN-50081-2
EN-50082-2
LVD:
EN-61010-1
Input: Choice of single or dual sensor input with
measurement choices of pH/ORP, conductivity/
resistivity, toroidal conductivity, flow, chlorine, dis-
solved oxygen, and dissolved ozone. Field-com-
missioned units allow user to change measure-
ments on either or both inputs. See combination
guide for valid combinations. For contacting con-
ductivity measurements, temperature element
must be a Pt 1000 RTD. For other measurements,
use either a Pt100 RTD, Pt1000 RTD, or 22k NTC
(D.O. only).
Outputs: Two 4-20 mA or 0-20 mA isolated outputs.
Continuously adjustable. Linear or logarithmic.
Maximum load 600 ohms. Output dampening with
time constant of 5 sec is user-selectable.
Alarms: Three alarm relays for process measure-
ment(s) or temperature. Alarm 3 can be
configured as a fault alarm, instead of a
process alarm. Each relay can be configured
independently. Alarm logic (high or low
activation or USP 23*) and deadband are
user-programmable. The USP 23* alarm can
be programmed to activate when the conduc-
tivity is within a user-selectable percentage of
the limit.
*conductivity/resistivity measurement only
Relays: Form C, single pole double throw, epoxy sealed
Weight/Shipping weight (rounded up to nearest lb or
nearest 0.5 kg): 3 lb (1.5 kg)/4 lb (2.0 kg)
Resistive
Inductive
28 Vdc
5.0 A
3.0 A
115 Vac
5.0 A
3.0 A
230 Vac
5.0 A
1.5 A