TENSPros
20
04
Alternate
program (A)
The stimulation of the CH2 will occur after the
1
st
contraction of CH1 is completed. In this
mode, the setting of ON time should be no less
than two times of the “Ramp” time. The OFF
time should be equal or more than the ON time.
ON time ≥Ramp up + Ramp down
OFF time ≥ON time
EMS WAVEFORMS
Constant (C)
Synchronous (S)
ON TIME
OFF TIME
CH1
CH2
Summary of Contents for Twin Stim Plus 3rd Edition
Page 1: ...T E N S P r o s Twin Stim Plus 3rd Edition Combo Stimulator INSTRUCTION MANUAL...
Page 2: ......
Page 39: ......