Fig 1.24 Digital Inputs
Caution
Applying the external voltage to digital inputs causes damage to the meter because DC 8V is
internally applied to
dry contacts
.
2 Digital Output Channels
Ch 1
Ch 2
Pin name DO
1
DO
2
Ground
DO
1G
DO
2G
*
DO
1G
and DO
2G
are different grounds according to each channel.
Item
Description
Connector
Screw Terminal[Pluggable]
Wire
1.25 to 3.5 mm2[ 24 to 14 AWG]
Signal type
Latch or Pulse
Max Load
AC/DC 400V 350mA[1000mA Peak]
Latch
On command
→
“on” outputs during the Pulse width time and returns to “off”
Off command
→
not available
Pulse
On command
→
“on” outputs
Off command
→
“off” outputs