![Ronde & Schwarz 1304.0004.03 User Manual Download Page 495](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985495.webp)
Data Management
R&S
®
FPL1000
495
User Manual 1178.3370.02 ─ 09
For the R&S
FPL1000, test reports are only available in Spectrum mode, and only if
the R&S
FPL1-K54 is installed.
For details on the provided information, see:
●
●
●
●
●
●
●
●
Result list
: a table that contains the trace values for each trace point
A maximum of 50001 measurement points per trace are written to the report. The
rest is dismissed.
If you have more measurement points, export the data to an ASCII file (see
●
Settings
: basic instrument and measurement-specific settings (as indicated in the
channel setup bar)
Select All / Select None
Selects or deselects all items in the corresponding area:
●
Diagrams
●
Tables
●
All items
Remote command:
8.5.2
Managing Templates
Access
: > "Report menu" > "Templates"
Some general functions to manage test report templates are available on all "Tem-
plates" subtabs.
Template name
Enter the path and filename of the template in the input field, or select the directory
using the "..." button.
Working with Test Reports