31
Auto Cal Menu
This menu is used to view and setup the Auto Calibration for the X96S Level with Density Compensation
Gage. It contains the following items:
ITEM
FUNCTION
Auto Cal Setup
Selecting this item takes the user to the
Auto Cal Setup
menu
Auto Cal Status
Selecting this item takes the user to the
Auto Cal Status
menu. The Auto Cal
Status shows the state for each Auto-Cal.
AC Filter
Shows and allows the user to set the
AC Filter
value. The
AC Filter
is the
time constant used for filtering the level detector signal during the auto calibration
G.Auto Delay
Shows and allows the user to set the
G.Auto Delay
value. G.Auto Delay is the
amount of time that must take place before the any of the other auto cal detectors can
begin an auto cal. This is to prevent simultaneous auto-calibrations.
Next Auto Time
Shows and allows the user to set the
Next Auto Time
value.
Next Auto Time
is the amount of time that must take place before the same auto-cal detector can
perform another auto-cal. This is to prevent the same detector from being in
continuous auto calibration.
Ref.Time
Shows and allows the user to set the
Ref.Time
value. The
Ref.Time
is the
amount of time it will take for an auto calibration to be completed. Recommend the
Ref.Time value should be at least 3 times the AC Filter value.
Auto Cal Setup Menu
ITEM
FUNCTION
Auto Cal 1
Selecting this item takes the user to the Auto Cal
1
configuration menu
Auto Cal 2
Selecting this item takes the user to the Auto Cal
2
configuration menu
Auto Cal 3
Selecting this item takes the user to the Auto Cal
3
configuration menu
Auto Cal 4
Selecting this item takes the user to the Auto Cal
4
configuration menu
Auto Cal 5
Selecting this item takes the user to the Auto Cal
5
configuration menu
Auto Cal 6
Selecting this item takes the user to the Auto Cal
6
configuration menu
Auto Cal 7
Selecting this item takes the user to the Auto Cal
7
configuration menu
Auto Cal 8
Selecting this item takes the user to the Auto Cal
8
configuration menu