GUI Reference
R&S
®
ZNB/ZNBT
296
User Manual 1173.9163.02 ─ 38
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Calibration Type
Selects the calibration type for the selected physical ports. For an overview refer to
Chapter 4.5.7.2, "Calibration Types"
on page 173. The calibration types PUOSM and
PTOSM are compatible with all port combinations.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Power Port
Selects the port for the scalar source power calibration.
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
. Next is unavailable (and a warning is displayed) if
the following happens:
●
Less than two ports have been selected
●
The selected characterization data do not cover all the ports to be calibrated.
●
The selected characterization data do not contain all standards needed for the
selected calibration type.
Step 2: Connections
Defines the port assignment(s) between the R&S
ZNB/ZNBT and the calibration unit.
Cal Softtool