Command Reference
R&S
®
ZNB/ZNBT
1147
User Manual 1173.9163.02 ─ 38
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Queries the warmup state of the connected calibration unit R&S ZV_Z5x. If several cal
units are USB-connected to the analyzer, the command queries the cal unit selected
via
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Possible responses are
1
(true, if the calibration unit has been connected for a suffi-
cient time to rach its operating temperature) or
0
(false).
0
is also returned if no calibra-
tion unit is connected.
Example:
See
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:CATalog?
Queries the numbers of all configured external generators. The response is a string
containing a comma-separated list of generator numbers.
Suffix:
<Gen>
.
Number of the configured generator. This suffix is ignored; the
command affects all generators.
Example:
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21'
SYST:COMM:RDEV:GEN2:DEF 'Ext. Gen. 2', 'SME02',
'gpib0', '22'
Configure two R&S SME02 generators as external generators
no. 1 and 2, assigning different names and GPIB addresses.
SYST:COMM:RDEV:GEN1:CAT?
Query the generator numbers. The response is '1, 2'.
SYST:COMM:RDEV:GEN1:COUN?
Query the number of configured generators. The response is 2.
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator:COUNt?
Queries the number of configured external generators. The response is an integer
number of generators.
Example:
See
SYSTem:COMMunicate:RDEVice:GENerator<Gen>: