Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
85
User Manual 1179.3315.02 ─ 01
INPut<ip>:RF:CAPMode:WAVeform:SRATe
........................................................................ 90
INSTrument:BLOCk:CHANnel[:SETTings]:SOURce<si>
.......................................................92
INSTrument:BLOCk:CHANnel[:SETTings]:SOURce<si>:CONFig
.......................................... 92
INSTrument:BLOCk:CHANnel[:SETTings]:SOURce<si>:TYPE
............................................. 93
SYSTem:COMMunicate:RDEVice:OSCilloscope[:STATe]
......................................................93
SYSTem:COMMunicate:RDEVice:OSCilloscope:TCPip
........................................................94
SYSTem:COMMunicate:RDEVice:OSCilloscope:PSMode[:STATe]
........................................ 94
SYSTem:COMMunicate:RDEVice:OSCilloscope:SRATe
.......................................................95
SYSTem:COMMunicate:RDEVice:OSCilloscope:VDEVice?
.................................................. 95
SYSTem:COMMunicate:RDEVice:OSCilloscope:VFIRmware?
..............................................95
INPut<ip>:ATTenuation:PROTection[:STATe]
<State>
This command turns the availability of attenuation levels of 10
dB or less on and off.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
ON | 1
Attenuation levels of 10
dB or less are not allowed to protect the
RF input connector of the connected instrument.
OFF | 0
Attenuation levels of 10
dB or less are not blocked. You must
provide appopriate protection for the RF input connector of the
connected instrument yourself.
*RST:
1
Example:
INP:ATT:PROT ON
Turns on the input protection.
INPut<ip>:ATTenuation:PROTection:RESet
[<DeviceName>]
This command resets the attenuator and reconnects the RF input with the input mixer
for the connected instrument after an overload condition occurred and the protection
mechanism intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status regis-
ter) and the
INPUT OVLD
message in the status bar are cleared.
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see the instrument's documentation.