Installing the module
R&S
®
TS-PMB
14
User Manual 1153.5233.12 ─ 06
5
Installing the module
The R&S TS-PMB is a module installed on the front panel of the base unit.
1. Install the module as described in the user manuals of the base units.
2.
WARNING!
Risk of electric shock. The test environment, e.g the UUT or additional
power supplies, can supply high voltages to the instruments. In this case, the volt-
age can also apply to the signal output connectors of the R&S
TSVP, in particular
the analog bus connector X2.
Therefore, do not connect or disconnect devices from the X2 connectors while con-
nected to an external power supply or UUT.
Always connect both ends of the cable connecting the R&S
CompactTSVP and
R&S
PowerTSVP. Thus, you avoid the risk of touching the X2 connector with a
possibly hazardous voltage applied.
Take the system into operation as described in the user manuals of the base units.
(starting with R&S
CompactTSVP with backplane version V4.x)
Matrix modules B R&S TS-PMB with change status V2.x (recognizable from the lack of
connector X1, last delivery in 2004) require a hardware change to TAZ 2.14 and a rear
IO module R&S
TS-PRIO plugged in to operated in the R&S
CompactTSVP with back-
plane version V4.x (starting with serial number 100109 on 09/2005). The 5-V power
supply and CAN bus are supplied via the R&S TS-PRIO.