Remote control commands
R&S
®
SMM100A
705
User Manual 1179.1341.02 ─ 05
The remote measurement is triggered with
). This command
also returns the measurement results. The local state is not affected, measurement
results can be retrieved with local state on or off.
Parameters:
<Continuous>
0 | 1 | OFF | ON
*RST:
0
Example:
:INIT1:CONT ON
Switches local state of continuous power measurement on.
Manual operation:
See
:READ<ch>[:POWer]?
Triggers power measurement and displays the results.
Note:
This command does not affect the local state, i.e. you can get results with local
state on or off. For long measurement times, we recommend that you use an SRQ for
command synchronization (MAV bit).
Suffix:
<ch>
.
1 to 3
Return values:
<Power>
float or float,float
The sensor returns the result in the unit set with command
Certain power sensors, such as the R&S NRP-Z81, return two
values, first the value of the average level and - separated by a
comma - the peak value.
Example:
:SENS1:UNIT DBM
Selects unit dBm for presentation of measurement result.
:READ1?
Queries the measurement result of the sensor.
-45.6246576745440230
-45.6 dBm were measured at the given frequency.
Example:
R&S NRP-Z81
:READ1?
-55.62403263352178,-22.419472478812476
-55.6 dBm is the measured average level, -22. 4 dBm is the
measured peak level at the given frequency.
Usage:
Query only
Manual operation:
See
"Level (Peak) / Level (Average)"
:SENSe<ch>:UNIT[:POWer]
<Power>
Selects the unit (Watt, dBm or dB
μ
V) of measurement result display, queried with
.
SENSe, READ, INITiate and SLISt subsystems