Multi-instrument setups
R&S
®
SMM100A
478
User Manual 1179.1341.02 ─ 05
See user manual R&S
SMM-K540, R&S
SMM-K541 Envelope Tracking and AM/AM,
AM/PM Predistortion.
9.6.1
General analog I/Q output settings
Access:
► Select "I/Q Analog > I/Q Analog Outputs > General".
For description of the impairments settings, see
Chapter 5.2, "Impairing the signal"
Settings:
State
Enables/disables the analog I/Q output.
Note:
By default, these output connectors are deactivated.
Remote command:
[:SOURce<hw>]:IQ:OUTPut:ANALog:STATe
Set to Default
Calls the default settings. The values of the main parameters are listed in the following
table.
Parameter
Value
"State"
Not affected by the "Set to Default"
"RF Envelope"
Off
"I/Q Output Type"
Depends on "System Configuration > External RF and I/Q > Preset behavior:
Keep connections to external instruments":
●
"Off":
Single Ended
●
"On":
Not affected by the "Set to Default"
"I/Q Level Vp (EMF)"
1 V
"Bias (EMF)"
0 mV
I/Q analog output settings