![Rohde & Schwarz SMM100A User Manual Download Page 480](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939480.webp)
Multi-instrument setups
R&S
®
SMM100A
480
User Manual 1179.1341.02 ─ 05
"Differential"
Option: R&S
SMM-K17
●
If "RF Envelope > Off"
The analog I/Q signal components are output at the I/Q and I/Q
Bar connectors.
●
If "RF Envelope > On"
The inverted envelope signal
Ē
is output at the I Bar connectors.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE
Mode
Selects the mode for setting the outputs, independent or together.
"Fixed"
The I/Q signal components settings are predefined fix values:
●
"Bias (EMF)" = "Offset (EMF)" = 0V
●
"I/Q Level Vp (EMF)"
–
Single ended mode: "I/Q Level Vp (EMF)" = 1 V
–
Differential mode: "I/Q Level Vp (EMF)" = 2 V
"Variable"
Option: R&S
SMM-K17
Enables modification of the I/Q signal components settings.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE
I/Q Level Vp EMF
In "Mode > Variable", sets the output voltage for the I and Q signal components.
The value range is adjusted so that the maximum overall output voltage does not
exceed 4V, see
"Maximum overall output voltage"
Remote command:
Couple I/Q Bias
In "Variable Mode", couples the bias setting of the I signal and the Q signal component.
With activated coupling, the settings of the I signal component automatically apply to
the Q signal component.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:COUPling[:STATe]
Bias (EMF)
If a bias is enabled, a DC voltage is superimposed upon the I or Q signal.
I/Q analog output settings