The dialog comprises the settings for selecting the analog output signal type (sin-
gle-ended or differential), enabling and generating an RF envelope signal, and add-
ing digital impairments to the signal.
5. To configure connected external instruments, use the "System Config > System
Configuration >
External RF and I/Q" dialog.
The parameters defining the voltage level of the output signal are interdependent.
Consider the limits listed in
"Maximum overall output voltage"
Maximum overall output voltage
The voltage of the analog output signals is defined as a combination of output voltage
of the I and Q signal components and an optional bias voltage. If the differential output
is used, an additional offset can be set, too. The values of these parameters are inter-
dependent so that the magnitude of the sum of output voltage and bias voltage must
not exceed 4V.
The following applies:
●
Single ended signal:
V
p
+ V
bias
≤
4
V
●
Differential signal:
0.5*V
p
+ |V
bias
| + 0.5*V
offset
≤
4
V
Where:
●
V
p
is the output voltage as set with the parameter
●
V
bias
is the bias voltage as set with the parameter
●
V
offset
is the offset between inverting and non-inverting output as set with the
For more information, see data sheet.