Multi-instrument setups
R&S
®
SMM100A
443
User Manual 1179.1341.02 ─ 05
● Select "Block Diagram > I/Q Stream Mapper > System Configuration > Multi
Instrument"
● On the instrument front panel, press [Setup].
Select "System Configuration > Multi Instrument".
For step-by-step description, see
Chapter 9.2.1.2, "How to connect and configure
instruments in primary-secondary instrument mode"
Settings
.............................................................................................. 443
.......................................................................................... 444
State
Activates the configuration.
With "State = Off", the instrument works in a standalone mode.
Remote command:
:SCONfiguration:MULTiinstrument:STATe
Multi Instrument Trigger
Sets the mode in that the instrument works.
To activate the mode, set "Multi Instrument > State = On".
"Primary"
Provides its trigger signal to the secondary instruments.
"Secondary"
Synchronizes to the supplied trigger signal from the primary instru-
ment.
Observe the "Synchronisation State" indication for information if the
synchronization signal is found and synchronization is achieved.
Remote command:
:SCONfiguration:MULTiinstrument:MODE
Synchronisation State
If "Multi Instrument Trigger = Secondary", indicates if synchronization is achieved.
"Sync"
Instrument is in sync.
It receives and decodes the synchronization and the trigger signal
from the primary instrument.