Remote control commands
R&S
®
SMM100A
716
User Manual 1179.1341.02 ─ 05
// Disconnect and disable remote control of the external instruments.
SCONfiguration:EXTernal:REMote:DISConnect[:ALL]
Example: Configuring primary-secondary mode
// For the
primary
instrument:
SOURce1:ROSCillator:SOURce INT
SCONfiguration:MULTiinstrument:MODe PRIMary
SCONfiguration:MULTiinstrument:STATe 1
SCONfiguration:APPLy
SOURce1:INPut:USER6:DIRection OUTP
OUTPut1:USER6:SIGNal INSinc
SOURce1:INPut:USER5:DIRection OUTP
OUTPut1:USER5:SIGNal INSinc
// For the
secondary
instrument:
SOURce1:ROSCillator:SOURce EXT
SOURce1:ROSCillator:EXTernal:FREQuency 10MHZ
SCONfiguration:MULTiinstrument:MODe SECondary
SCONfiguration:MULTiinstrument:STATe 1
SOURce1:INPut:USER5:DIRection INP
SOURce1:INPut:USER5:SIGNal OUTSinc
SOURce1:INPut:USER6:DIRection OUTP
OUTPut1:USER6:SIGNal INSinc
SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
// SYNC
Multi instrument configuration
............................................................................... 716
...............................................................................................717
External RF and I/Q instruments
...........................................................................718
13.15.1
Multi instrument configuration
The following commands configure multi-instrument settings.
Commands:
:SCONfiguration:MULTiinstrument:MODE
.........................................................................716
:SCONfiguration:MULTiinstrument:STATe
......................................................................... 717
:SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
...........................................717
:SCONfiguration:MULTiinstrument:MODE
<MSMode>
Sets if the instrument works as a primary or as a secondary instrument.
Parameters:
<MSMode>
PRIMary | SECondary
*RST:
PRIMary
SCONfiguration subsystem