Multi-instrument setups
R&S
®
SMM100A
444
User Manual 1179.1341.02 ─ 05
"No Sync"
Synchronization is not achieved, because:
●
The baseband generators in the secondary instruments or in the
primary instruments are not activated, see
●
There is no physical connection (the synchronization signal is not
connected as required).
For details, see
Chapter 9.2.1.2, "How to connect and configure
instruments in primary-secondary instrument mode"
●
The secondary instrument does not detect the synchronization
signal.
Remote command:
:SCONfiguration:MULTiinstrument:TRIGger:SYNChronization?
Global Connector Settings
Provides a quick access to the related global connector settings.
With "State = On", the synchronization signal is assigned per default as follows:
●
"Multi Instrument Trigger > Primary"
"Connector = User 6": "Direction = Output", "Signal = Baseband Sync Out"
●
"Multi Instrument Trigger > Secondary"
"Connector = User 6": "Direction = Output", "Signal = Baseband Sync Out"
"Connector = User 5": "Direction = Input", "Signal = Baseband Sync In"
See also:
●
Chapter 9.2.1.2, "How to connect and configure instruments in primary-secondary
instrument mode"
See also
Chapter 11.2, "Configuring local and global connectors"