Configuring the internal baseband source
R&S
®
SMM100A
110
User Manual 1179.1341.02 ─ 05
Refer to
"Impact of the trigger modes on the signal generation"
on page 96 for a
detailed description of the provided trigger modes.
Remote command:
[:SOURce<hw>]:BB:DM[:TRIGger]:SEQuence
[:SOURce<hw>]:BB:ARBitrary[:TRIGger]:SEQuence
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence
Signal Duration
Enters the length of the signal sequence to be output in the "Single" trigger mode.
Use this parameter to output part of the signal deliberately, an exact sequence of the
signal, or a defined number of repetitions of the signal.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:SLENgth
[:SOURce<hw>]:BB:ARBitrary:TRIGger:SLUNit
[:SOURce<hw>]:BB:MCCW:TRIGger:SLENgth
Running/Stopped
With enabled modulation, displays the status of signal generation for all trigger modes.
●
"Running"
The signal is generated; a trigger was (internally or externally) initiated in triggered
mode.
●
"Stopped"
The signal is not generated and the instrument waits for a trigger event.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:RMODe?
[:SOURce<hw>]:BB:ARBitrary:TRIGger:RMODe?
[:SOURce<hw>]:BB:MCCW:TRIGger:RMODe?
Arm
Stops the signal generation until subsequent trigger event occurs.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:ARM:EXECute
[:SOURce<hw>]:BB:ARBitrary:TRIGger:ARM:EXECute
[:SOURce<hw>]:BB:MCCW:TRIGger:ARM:EXECute
Execute Trigger
For internal trigger source, executes trigger manually.
Remote command:
[:SOURce<hw>]:BB:DM:TRIGger:EXECute
[:SOURce<hw>]:BB:ARBitrary:TRIGger:EXECute
[:SOURce<hw>]:BB:MCCW:TRIGger:EXECute
Source
The following sources of the trigger signal are available:
●
"Internal"
The trigger event is executed manually by the "Execute Trigger".
●
"External Global Trigger"
Common functions and settings in the baseband domain