Remote Control Commands
R&S
®
SMF100A
408
Operating Manual 1167.2319.02 ─ 12
Manual operation:
See
" Auto Scale - Power Analysis "
:SENSe[:POWer]:SWEep:FREQuency:YSCale:AUTO:RESet
Resets the Y scale to suitable values after the use of auto scaling in the expanding
mode. For this mode, the scale might get expanded because of temporarily high power
values. The reset function resets the diagram in such a way that it matches smaller
power values again.
Example:
SENS:SWE:FREQ:YSC:AUTO:RES
resets auto scale
Usage:
Event
Options:
Option R&S SMF-K28
Manual operation:
See
" Reset Auto Scale - Power Analysis "
:SENSe[:POWer]:SWEep:FREQuency:YSCale:MAXimum
<Maximum>
Sets the maximum value for the y axis of the measurement diagram.
Parameters:
<Maximum>
float
Range:
-200 to 100
Increment: 0.01
*RST:
40
Example:
SENS:SWE:FREQ:YSC:MAX 10DBM
sets 10 dBm as the upper limit of the measurement diagram.
Options:
Option R&S SMF-K28
Manual operation:
See
" Min - Max y-Axis - Power Analysis "
:SENSe[:POWer]:SWEep:FREQuency:YSCale:MINimum
<Minimum>
Sets the minimum value for the y axis of the measurement diagram.
Parameters:
<Minimum>
float
Range:
-200 to 100
Increment: 0.01
*RST:
-40
Example:
SENS:SWE:FREQ:YSC:MIN -10DBM
sets -10 dBm as the lower limit of the measurement diagram.
Options:
Option R&S SMF-K28
Manual operation:
See
" Min - Max y-Axis - Power Analysis "
Power Sensor Measurement Subsystems