Remote Control Commands
R&S
®
SMC100A
211
Operating Manual 1411.4060.32 ─ 04
Example:
SYST:FPR
all instrument settings (also those that are not currently active) are
reset to the factory values.
Usage:
Event
7.3 CALibration Subsystem
The CALibration system contains the commands for adjustment. Adjustment is triggered
by the query commands. The response "0" indicates error-free adjustment, and the
response "1" means that an error occurred during adjustment.
...........................................................................................211
...........................................................................211
.......................................................................211
................................................................................212
............................................................................212
:CALibration:ALL[:MEASure]?
The command starts all internal adjustments for which no external measuring equipment
is needed.
Return values:
<Measure>
0 | 1
Example:
CAL:ALL:MEAS?
starts the adjustment of all functions for the entire instrument.
Response: "0"
adjustment has been performed successfully.
Usage:
Query only
SCPI confirmed
:CALibration<hw>:FMOFfset[:MEASure]?
The command starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure>
0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage:
Query only
:CALibration<hw>:FREQuency[:MEASure]?
The command starts all adjustments which affect the frequency.
CALibration Subsystem