Remote Control Commands
R&S
®
SMC100A
212
Operating Manual 1411.4060.32 ─ 04
Return values:
<Measure>
0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage:
Query only
:CALibration<hw>:LEVel[:MEASure]?
The command starts all adjustments which affect the level.
The acquired correction values improve the settling time and the signal quality.
Return values:
<Measure>
0 | 1
Example:
CAL:LEV:MEAS?
starts adjustments for maximum level accuracy.
Response: "0"
adjustment has been performed successfully.
Usage:
Query only
:CALibration<hw>:LEVel:EXTern:DATA
<Data>
Selects the data used for of external level correction.
By default the instrument uses correction data obtained in the factory before delivery. In
addition, customer data can be used for external level correction. The customer data is
obtained using a R&S NRP power sensor. External level correction is a protected function
(see service manual, chapter 2, "Adjustment").
Parameters:
<Data>
FACT | CUST
*RST:
FACTory
Example:
CAL:LEV:EXT:DATA FACT
selects the use of the data aquired at the factory for external level
correction.
7.4 DIAGnostic Subsystem
The
DIAGnostic
system contains the commands used for instrument diagnosis and
servicing. SCPI does not define any DIAGnostic commands; the commands listed here
are all Device-specific. All DIAGnostic commands are query commands which are not
influenced by
................................................................................................213
..................................................................................213
DIAGnostic Subsystem