Instrument Function
R&S
®
SMC100A
140
Operating Manual 1411.4060.32 ─ 04
"Triangle"
One sweep runs from start to stop frequency and back, i.e. the shape
of the sweep resembles a triangle. Each subsequent sweep starts at
the start frequency.
SCPI command:
Step Lin/Log - Frequency Sweep
Sets the step width for the individual sweep steps. This entry is effective for all sweep
modes.
"Step Lin" or "Step Log" is displayed depending on whether "Spacing Lin" or "Log" is
selected.
"Step Lin"
With the linear sweep, the step width is a fixed frequency value which
is added to the current frequency. The linear step width is entered in
Hz.
SCPI command:
"Step Log"
With the logarithmic sweep, the step width is a constant fraction of the
current frequency. This fraction is added to the current frequency. The
logarithmic step width is entered in %.
SCPI command:
Dwell Time - Frequency Sweep
Sets the dwell time. The dwell time determines the duration of the individual sweep steps.
The "Dwell Time" set by the user is used as the step time of the sweep. The effective net
dwell time is shorter, reduced by the setting time. This setting time may be greater than
the time specified in the data sheet.
Note:
It is recommended to switch off the GUI update for optimum sweep performance espe-
cially with short dwell times (see
SCPI command:
RF Signal - RF Block