Remote Control Commands
R&S
®
SMC100A
293
Operating Manual 1411.4060.32 ─ 04
Example:
FREQ:STAR
sets the start frequency to 100 MHz.
FREQ:STOP
sets the stop frequency to 500 MHz.
SWE:SPAC LIN
sets linear sweep spacing.
SWE:POIN 401
sets 401 sweep steps for linear sweep spacing. The sweep step
width (
STEP
) is automatically set to 1 MHz.
[:SOURce<hw>]:SWEep[:FREQuency]:SHAPe
<Shape>
The command sets the cycle mode for a sweep sequence (shape).
Parameters:
<Shape>
SAWTooth | TRIangle
SAWTooth
One sweep runs from start to stop frequency. Each subsequent
sweep starts at the start frequency, i.e. the shape of the sweep
sequence resembles a sawtooth.
TRIangle
One sweep runs from start to stop frequency and back, i.e. the
shape of the sweep resembles a triangle. Each subsequent sweep
starts at the start frequency.
*RST:
SAWTooth
Example:
SOUR:SWE:SHAP TRI
selects the sweep cycle with alternating ascending and descend-
ing sweep directions.
[:SOURce<hw>]:SWEep[:FREQuency]:SPACing
<Spacing>
The command selects the sweep spacing.
The keyword
[:FREQuency]
can be omitted (see example). The command is then
SCPI-compliant.
SOURce Subsystem