Trying Out the Instrument
R&S
®
RTM2000
34
Getting Started 1317.4710.02 ─ 0101
7. Press "Measure 2" until "On" is highlighted.
8. Press "Meas. Type" until "Base Level" is highlighted.
9. Press "Source" until "CH1" is highlighted.
The top and base level values of the active signal are displayed in the result table.
10. Press QUICK MEAS to turn off the quick measurement mode.
4.5 Performing Cursor Measurements
While the quick and automatic measurements provide results for the entire (visible)
waveform or at pre-defined points, a cursor measurement allows you to specify at
which points in the diagram you require results. Depending on the type of measure-
ment, two or three cursors are positioned on the screen and the results for those points
are displayed.
For example, using the voltage & time cursor measurement, you can position two verti-
cal cursors at the pulse edges and the horizontal cursors at the levels. Thus the pulse
width as well as top and base level voltages are measured.
Performing Cursor Measurements