Preparing for Use
R&S
®
RTM2000
8
Getting Started 1317.4710.02 ─ 0101
2 Preparing for Use
This section describes the basic steps to be taken when setting up the R&S
RTM for
the first time.
Risk of instrument damage
Note that the general safety instructions also contain information on operating condi-
tions that will prevent damage to the instrument. The instrument's data sheet may con-
tain additional operating conditions.
Risk of instrument damage during operation
An unsuitable operating site or test setup can cause damage to the instrument and to
connected devices. Ensure the following operating conditions before you switch on the
instrument:
●
All fan openings are unobstructed and the airflow perforations are unimpeded. The
minimum distance from the wall is 10 cm.
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are correctly connected and are not overloaded.
2.1 Unpacking and Checking the Instrument
To remove the instrument from its packaging and check the equipment for complete-
ness, proceed as follows:
1. Pull off the polyethylene protection pads from the instrument's rear feet and then
carefully remove the pads from the instrument handles at the front.
2. Pull off the corrugated cardboard cover that protects the rear of the instrument.
3. Carefully unthread the corrugated cardboard cover at the front that protects the
instrument handles and remove it.
4. Check the equipment for completeness using the delivery note and the accessory
lists for the various items.
Unpacking and Checking the Instrument