Instrument Tour
R&S
®
RTM2000
19
Getting Started 1317.4710.02 ─ 0101
LEVEL
The rotary knob changes the trigger treshold voltage for all trigger types that require
one trigger level. Some trigger types require two trigger levels, for example, the rise
time/fall time trigger. These levels can be set in the trigger setup menu.
Turn clockwise to move the trigger level up. Press the knob to set the level to 50% of
the signal amplitude.
If a B-trigger is enabled, the knob can set the level for both the A- and B-trigger. To
assign the level to the A- or B-trigger, use "Trigger Level" in the "Trigger" menu.
SETUP
Opens the "Trigger" menu.
SOURCE
Opens the "Trigger Source" menu for the A-trigger. Press the key repeatedly until the
required source is selected. The key lights up in the color of the selected trigger chan-
nel. The selected source is shown in the information bar.
SLOPE
If you have selected "Edge" trigger as trigger type with an analog trigger source, the
SLOPE key toggles the trigger slope. The current setting is shown by an icon in the
information bar.
If a B-trigger is enabled, the key can set the slope for both the A- and B-trigger. To
assign the slope to the A- or B-trigger, use "Trigger Level" in the "Trigger" menu.
3.1.6 VERTICAL Controls
The keys and knobs in the VERTICAL functional block select a signal and adjust the
vertical scale and position of the selected waveform.
Front Panel