![Rohde & Schwarz RT-ZS60 User Manual Download Page 67](http://html.mh-extra.com/html/rohde-and-schwarz/rt-zs60/rt-zs60_user-manual_1477877067.webp)
Functional Check
R&S
®
RT-ZS60
45
User Manual 1418.7342.02 ─ 01
6.3
DC Accuracy
The gain of the probe-oscilloscope system and the accuracy of the integrated DC
voltmeter are checked.
Test setup:
●
Probe connected to CH1 of the R&S RTO oscilloscope
●
INPUT HI/LO of the DMM connected to OUTPUT HI/LO of the DC
voltage source by banana leads
●
BNC to 4 mm dual banana adapter plugged onto the banana leads
on the DMM – ensure that BNC shield is connected to LO
●
Signal and ground terminal of the probe connected by mini clips to
the BNC to 4 mm dual banana adapter; make sure that probe ground
is connected to BNC shield
Lead connected to the input socket using a signal adapter, square pin
R&S RTO settings:
●
Horizontal scale = 10 µs / div
●
Vertical scale = 1.6 V / div
●
Trigger source = Ch1, Trigger mode = Auto, Run cont
●
ACQUISITION key > Decimation = 'High res' > Waveform Arithmetics
= 'Average' > Average Count = 16
●
"Meas" menu > Setup > Source = 'Ch1' > Main Measurement = 'Mean'
> State = 'on'
●
CH1 key > Offset = 0 V, Position = 0 div
●
"Vertical" menu > Probe Setup > Ch1 > ProbeMeter = 'on'
DC Accuracy