Putting into Operation
R&S
®
RT-ZS60
21
User Manual 1418.7342.02 ─ 01
To assign "Set offset to mean" to the micro button:
► "Vertical" menu > "Probe Setup" > "Ch" tab > "Micro button action" = "Set offset
to mean"
The maximum input voltage is ±30 V. A higher input voltage may destroy the
probe.
2.4
R&S ProbeMeter
The integrated voltmeter can measure DC voltages with very high precision com-
pared to the oscilloscope's DC accuracy. The DC measurement is performed con-
tinuously and in parallel to the time domain measurement of the oscilloscope. High-
precision measurements are achieved through immediate digitization of the mea-
sured DC voltage inside the probe.
To activate the R&S ProbeMeter:
► Vertical > Probe Setup > Ch > R&S ProbeMeter 'on'
After the R&S ProbeMeter has been activated, the measured values are displayed
on the screen of the oscilloscope.
Advantages of the R&S
ProbeMeter:
●
Enables the user to check DC voltages with different levels without having to
adjust the measurement range of the oscilloscope.
●
Provides a simple means of setting the oscilloscope's trigger level and vertical
scaling if a waveform is not visible.
●
Independent of oscilloscope settings for offset, position, vertical scale, horizontal
scale, and trigger.
●
True DC measurement (integration time > 100 ms), not mathematical average
of displayed waveform.
●
Measurement range ±8 V + offset compensation setting. Maximum measure-
ment accuracy is achieved when offset compensation is switched off.
●
High measurement accuracy and low temperature sensitivity, typically 0.1% of
reading over the entire temperature range.
R&S ProbeMeter