![Rohde & Schwarz RT-ZS60 User Manual Download Page 54](http://html.mh-extra.com/html/rohde-and-schwarz/rt-zs60/rt-zs60_user-manual_1477877054.webp)
Measurement Principles
R&S
®
RT-ZS60
32
User Manual 1418.7342.02 ─ 01
4.1.2
Connection Inductance
The connection inductance L
con
is caused by connecting the probe to the DUT. In
contrast to the probe-specific bandwidth, the connection inductance mainly
depends on the type of connection that the user selects.
The connection inductance
●
Increases with the length of the connection and the size of the resulting loop
area A.
●
Reduces the usable bandwidth and causes ringing with signals having a short
rise time (due a series resonance with the input capacitance).
●
Should be as small as possible (short lead length) in order to maintain high signal
integrity.
●
Long leads on the signal input are especially problematic.
con
in
con
resonance
con
L
BW
C
L
f
A
L
1
to
al
proportion
2
1
≈
to
al
proportion
Fig. 4-4: Ground connection and connection inductance
Signal Integrity of the Transferred Signal