Putting into Operation
R&S
®
RT-ZC20
19
User Manual 1409.7814.02 ─ 02
7. If zero adjustment is not possible, turn the coarse adjustment trimmer to bring
the trace within the range of adjustment by the zero adjustment dial, see
chapter 2.4, "Description of the Probe"
Note:
Use the trimmer carefully to prevent damaging it. Use a flathead screw-
driver that is made of non-conductive material like ceramics. Recommended
tip shape: thickness 4
mm, width 1.8
mm, length 10
mm or longer.
Risk of instrument damage
When opening the sensor head of the probe, be sure to operate with the
opening lever. If an upper core is forced to open when the sensor head is
locked, the open-close mechanism can be damaged.
3.4
Connecting the Probe to the DUT
1. Check that the system is safe, and that the preparations described in the pre-
ceding sections have been carried out.
2. Pull the sensor slider, so that the clamp opens.
3. Align the sensor so that the current direction indication matches the direction
of current flow through the conductor to be measured. The conductor should
be in the center of the clamp aperture because the measurement may be
affected by the position within the clamp aperture of the conductor being mea-
sured.
Connecting the Probe to the DUT