Remote Control - Command Description
R&S
®
UPP
732
User Manual 1411.1055.32 ─ 10
WAIT
CONTinous
measurement in which the
SENSe<n1>:DATA<n2>
commands, which are used to query the measurement results,
stop the IEC/IEEE bus until a measurement result is available.
RSTart
Starts a single measurement or a sweep. At the same time
resets the measurement traces and restart the Play function.
Manual operation:
See
See
See
See
See
See
See
See
See
See
See
See
See
See
INITiate:CONTinuous:TIMeout
<timeout>
Only in conjunction with the command
INITiate:CONTinuous WAIT
. Defines the
maximum time the bus is stopped.
If timeout occurs without a new measurement result being available, the last measure-
ment result that was read out is returned.
Parameters:
<timeout>
0 … 1000 s
INITiate:FORCe
<force>
Parameters:
<force>
STARt | STOP | SINGle | CONTinuous
STARt
Start of a continuous measurement.
Identical to the [Start] stroke.
Identical to
INITiate:CONTinuous ON
; however, without
changing the
INITiate
presetting.
STOP
Stops a measurement as if [STOP/CONT] was pressed while a
measurement was in progress. (Identical to the
ABORt
com-
mand).