Termination event
… stops the current sweep immediately and causes the continuous sweep to be restar-
ted (see
Chapter 5.20.1, "Overview of Sweep Options"
Z stop value reached
The continuous sweep is stopped; the sweep system changes to the "Sweep Waiting"
state (only with 2-dimensional generator sweeps, see
Chapter 5.9.3, "Two-Dimensional
5.20.3.2
Single sweep
Single sweeps are started by pressing the [SINGLE] key and are performed once only.
When the stop value (of the x-axis) has been reached, the sweep system switches to
the "Sweep Waiting" state. With a 2-dimensional sweep, every new Z point (and the
associated sweep over the x-axis) must therefore be retriggered.
The sweep state is indicated by "Sweep Single ".
A current single sweep is interrupted or terminated by the following events:
START
Pressing [START ]
●
interrupts the currently running sweep at once
●
deletes the sweep curves (in multiscan mode all scans are deleted) including the
min/max curves
●
resets the Min/Max. data and the limit violations of the sweep curves
●
starts a continuous sweep
Remote command:
ON
STARt
SINGLE
Pressing [SINGLE ] stops the current sweep immediately and causes the single sweep
to be restarted.
Remote command:
OFF
SINGle
STOP/CONT
Pressing [STOP/CONT ] terminates the current measurement and pauses the sweep
system (Sweep Stopped state).
The single sweep is reactivated either using [SINGLE ] (sweep starts from the begin-
ning) or by pressing [STOP/CONT ] again (sweep continues from the point at which it
was stopped).
Remote command:
STOP
Sweep Options
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...