CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:RESult?
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:SELect
<MeasType>
This command selects a power measurement and turns the measurement on.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<MeasType>
MCACpower | ACPower | CPOWer | OBANdwidth | OBWidth |
TDMA
MCACpower | ACPower
Adjacent Channel Leakage Ratio measurement (ACLR) aka
adjacent channel power or multi carrier adjacent channel.
CPOWer
Chanel power measurement.
OBANdwidth | OBWidth
Occupied Bandwidth measurement.
TDMA
TDMA power measurement.
Example:
CALC:MARK:FUNC:POW:SEL CPOW
Selects the channel power measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]
<State>
This command turns a power measurement on and off.
You can select a power measurement with
CALCulate<1...2>:MARKer<1...6>:
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<State>
ON | OFF
ON
Turns the power measurement on.
OFF
Performs a basic spectrum measurement.
*RST:
OFF
Example:
CALC:MARK:FUNC:POW OFF
Switches off the power measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:PRESet
<Standard>
This command selects one of the predefined configurations for a telecommunications
standard. This command only works if you have turned on power measurements with
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer[:STATe]