Instrument Functions
R&S
®
NGA100
45
User Manual 5601.8919.02 ─ 03
3. Set the required channel fusion mode:
● "Off": Channel operates separately
● "Series": In series mode, the two terminals in the middle are used for output.
There is an internal series link activated to connect channel 1 and channel 2
output. If this series mode is set, a "SER" indicator is displayed on the channel
status bar. For details, see
Figure 6-4: Series mode setting and output of one-channel display
Note: The "Digital IO", "EasyArb" and "EasyRamp" functions are disabled in
this mode.
● "Parallel": In parallel mode, an external wire is required to connect the "CH1"
and "CH2" terminals. All channels can be used as output. If parallel mode is
set, a "PAR" indicator is displayed on the channel status bar. See
Channel Fusion