Remote control commands
R&S
®
NGP800
141
User Manual 5601.5610.02 ─ 09
[SOURce:]VOLTage:PROTection[:STATe]
......................................................................... 141
[SOURce:]VOLTage:PROTection:CLEar
........................................................................... 141
[SOURce:]VOLTage:PROTection:LEVel
............................................................................ 141
[SOURce:]VOLTage:PROTection:TRIPped?
......................................................................142
[SOURce:]VOLTage:PROTection[:STATe]
<En/Disable volt protection>[, <Channel
list>]
[SOURce:]VOLTage:PROTection[:STATe]?
[<Channel list>]
Sets or queries the OVP state of the previous selected channel.
Parameters:
<state>
0
OVP is deactivated
1
OVP is activated
Parameters for setting and query:
<Channel list>
<list>
Example:
VOLT:PROT? (@1)
Queries OVP state at channel 1.
Example:
See
Example "Configuring the overvoltage protection"
[SOURce:]VOLTage:PROTection:CLEar
[<Channel list>]
[SOURce:]VOLTage:PROTection:CLEar
[<Channel list>]
Resets the OVP state of the selected channel. If an OVP event has occurred before,
the reset also erases the message on the display.
Parameters:
<Channel list>
<list>
Example:
VOLT:PROT:CLEAR (@1)
Resets OVP state at channel 1.
Example:
See
Example "Configuring the overvoltage protection"
Usage:
Setting only
[SOURce:]VOLTage:PROTection:LEVel
[<New value for voltage protection>,
<Channel list>]
[SOURce:]VOLTage:PROTection:LEVel?
[<Channel list>]
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<voltage>
<numeric value> | MIN | MINimum | MAX | MAXimum | DEF |
DEFault
Configuration commands