Sub Block Count
Defines the number of sub blocks. By default, the familiar SEM measurement with just
one single block of ranges is configured.
Remote command:
Sub Block Center Frequency
Defines the center frequency for an individual sub block. The center frequency deter-
mines the reference range used for each block.
For measurements with only 1 sub block, this setting corresponds to the global setting
in the "Frequency" settings (see
Remote command:
Standard / MSR Settings
Defines the use of a standard settings file or a multi-standard radio configuration for a
particular sub block. For details see
chapter 5.6.5.6, "Standard Files"
chapter 5.6.5.5, "MSR Settings"
Note that
either
a standard
or
an MSR setting can be selected; if one is selected, the
other is disabled.
Remote command:
chapter 11.5.6.3, "Configuring a Multi-SEM Measurement"
Edit sweep list
Switches to the "Sweep List" tab of the "Spectrum Emission Mask" dialog box to con-
figure the individual frequency ranges and mask limits for the corresponding sub block.
See
5.6.5.3
Reference Range
The range centered around the center frequency is defined as the reference range for
all other ranges in the sweep list.
In the "Reference Range" tab of the "Spectrum Emission Mask" dialog box you define
the general settings for the reference range.
Spectrum Emission Mask (SEM) Measurement