![Rohde & Schwarz FSW13 User Manual Download Page 388](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847388.webp)
Common Measurement Settings
R&S
®
FSW
388
User Manual 1173.9411.02 ─ 19
...............................................................................................389
.........................................................................................390
....................................................................................391
...................................................................................... 391
....................................................................................... 391
........................................................................................391
Center frequency
Defines the normal center frequency of the signal. The allowed range of values for the
center frequency depends on the frequency span.
span > 0: span
min
/2 ≤ f
center
≤ f
max
– span
min
/2
zero span: 0 Hz ≤ f
center
≤ f
max
f
max
and span
min
are specified in the data sheet.
Remote command:
Span
Defines the frequency span. The center frequency is kept constant. The following
range is allowed:
Frequency and Span Configuration