Remote Commands
R&S
®
FPL1000
743
User Manual 1178.3370.02 ─ 03
Example:
CALC:MARK:FUNC:FPE:STAT ON
Activates marker peak search
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X?
This command queries the position of the peaks on the x-axis.
The order depends on the sort order that has been set with
MARKer<m>:FUNCtion:FPEaks:SORT
.
Suffix:
<n>, <m>
.
irrelevant
Return values:
<PeakPosition>
Position of the peaks on the x-axis. The unit depends on the
measurement.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y?
This command queries the position of the peaks on the y-axis.
The order depends on the sort order that has been set with
MARKer<m>:FUNCtion:FPEaks:SORT
.
Suffix:
<n>, <m>
.
irrelevant
Return values:
<PeakPosition>
Position of the peaks on the y-axis. The unit depends on the
measurement.
Usage:
Query only
8.8.8.9
Noise Measurement Marker
The following commands control the noise measurement marker function.
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
.....................................................743
CALCulate<n>:MARKer<m>:FUNCtion:NOISe[:STATe]
.....................................................744
CALCulate<n>:MARKer<m>:FUNCtion:NOISe:RESult?
This command queries the result of the noise measurement.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Setting Basic Spectrum RF Measurement Parameters