Remote Commands
R&S
®
FPL1000
669
User Manual 1178.3370.02 ─ 03
Parameters:
<Type>
LEVel
The trigger event for the gate to open is a particular power level.
After the gate signal has been detected, the gate remains open
until the signal disappears.
Note:
If you perform gated measurements in combination with
the IF Power trigger, the R&S
FPL1000 ignores the holding time
for frequency sweep, FFT sweep, zero span and I/Q mode mea-
surements.
EDGE
The trigger event for the gate to open is the detection of the sig-
nal edge.
After the gate signal has been detected, the gate remains open
until the gate length is over.
*RST:
EDGE
Example:
SWE:EGAT:TYPE EDGE
Manual operation:
See
8.8.5 Configuring the Data Input and Output
The following commands are required to configure data input and output.
................................................................................ 671
........................................................................................ 681
8.8.5.1
RF Input
INPut:ATTenuation:PROTection:RESet
............................................................................669
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occurred and the protection mechanism intervened. The
error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
(See
STATus:QUEStionable:POWer[:EVENt]?
Tus:QUEStionable:POWer Register"
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see
Usage:
Event
Setting Basic Spectrum RF Measurement Parameters