Remote Commands
R&S
®
FPL1000
681
User Manual 1178.3370.02 ─ 03
Parameters:
<Level>
-20 to +20 dBm
Range:
-20 dBm to 20 dBm
*RST:
-10 dBm
Example:
PMET2:TRIG:LEV -10 dBm
Sets the level of the trigger
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger:SLOPe
<Edge>
This command selects the trigger condition for external power triggers.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Edge>
POSitive
The measurement starts in case the trigger signal shows a posi-
tive edge.
NEGative
The measurement starts in case the trigger signal shows a neg-
ative edge.
*RST:
POSitive
Example:
PMET2:TRIG:SLOP NEG
Manual operation:
See
[SENSe:]PMETer<p>:TRIGger[:STATe]
<State>
This command turns the external power trigger on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
PMET2:TRIG ON
Switches the external power trigger on
Manual operation:
See
" Using the power sensor as an external trigger "
8.8.5.3
Configuring the Outputs
The following commands are required to provide output from the R&S
FPL1000.
Setting Basic Spectrum RF Measurement Parameters