Remote Commands
R&S
®
FPL1000
670
User Manual 1178.3370.02 ─ 03
INPut:IMPedance
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Example:
INP:IMP 75
Manual operation:
See
See
INPut:SELect
<Source>
This command selects the signal source for measurements, i.e. it defines which con-
nector is used to input data to the R&S
FPL1000.
Parameters:
<Source>
RF
Radio Frequency ("RF INPUT" connector)
*RST:
RF
Manual operation:
See
INPut:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FPL1000.
See the R&S
FPL1000 Getting Started manual for details.
Parameters:
<State>
ON | 1
User port is switched to INPut
OFF | 0
User port is switched to OUTPut
*RST:
1
INPut:UPORt[:VALue]?
This command queries the control lines of the user ports.
For details see
Return values:
<Value>
bit values in hexadecimal format
TTL type voltage levels (max. 5V)
Range:
#B00000000 to #B00111111
Setting Basic Spectrum RF Measurement Parameters